Vue de la page 236
Reference Manual
00809-0100-4530, Rev BA
July 2009
B-15
Rosemount 5300 Series
Figure B-15. Approval Label
Canadian Standards Association
(CSA) Foundation Fieldbus
model
I6 Intrinsically Safe Ex ia.
Class I, Division 1, Groups A, B, C and D.
Temperature code T4.
F
OUNDATION Fieldbus model: U
i
=30 Vdc, I
i
=300 mA, P
i
=1.3 W,
C
i
=0 nF, L
i
=0 H.
Control Drawing: 9240 030-937.
Ambient temperature limits: -50
°C to + 60 °C
Figure B-16. Approval Label
Canadian Standards Association
(CSA) FISCO model
IF Intrinsically Safe Ex ia.
Class I, Division 1, Groups A, B, C and D.
Temperature code T4.
FISCO model: U
i
=17.5 Vdc, I
i
=380 mA, P
i
=5.32 W,
C
i
=0 nF, L
i
=0 H.
Control Drawing: 9240 030-937.
Ambient temperature limits: -50
°C to + 60 °C