Reference Manual
00809-0100-4530, Rev BA
July 2009
7-43
Rosemount 5300 Series
LED ERROR MESSAGES For Rosemount 5300 transmitters without display, a flashing Light Emitting
Diode (LED) is used for presentation of error messages.
Figure 7-25. Rosemount 5300
transmitters without display use
a LED for presentation of error
messages.
In normal operation, the LED flashes once every other second. When an error
occurs, the LED flashes a sequence that corresponds to the Code number
followed by a five second pause. This sequence is continuously repeated.
The following errors can be displayed:
Table 7-12. LED error codes.
Example
Modem error (code 6) is displayed as the following flash sequence:
3
Flashing LED
Code Error
0
Ram Failure
1
FPROM
2
HREG
4
Microwave Module
5
Display
6
Modem
7
Analog Out
8
Internal Temperature
11
Hardware
12
Measurement
14
Configuration
15
Software
Seconds